| Ridgelimeter | to Top |
Directions:
Measurement of gel rigidity (Limits of sag: 10 to 340/o)
| 1. | Material to be tested must completely fill a Ridgelimeter glass. A collar of masking tape will permit an over fill and gel can be leveled with wire slicer before measuring. |
| 2. | Material must be removed from Ridgelimeter glass without being damaged. For a gelatin gel the gel can be removed by dipping the glass into warm water until the gel loosens from the glass. For pectin, starch and egg gels, the gels can be removed by air displacement. This is done by inserting a knife blade between:the side of the glass and the gel. At this point of knife entry air is displaced. By rotating the glass the air will rotate around the gel-and thereby loosen the gel from the glass. |
| 3. | Quickly slip the gel from the glass onto a watch plate. After 60 seconds, take the reading. |
| 4. | To take a reading the black knob is turned until the tip of the pointer touches the surface of the gel. |
| 5. | To read the measurement the vertical column gives the whole number. The number is read at the point where the bottom of the dial meets the vertical column. The dial which rotates horizontally gives the tenth of a number. The total reading, as indicates the percent sag.
|